UD TECHNOLOGY
PROBLEMS WITH PEAK SHAPE
Poor peak shape can cause a loss of resolution between adjoining peaks, it can lead to poor injection to injection reproducibility, or it may cause an injection to fail method acceptance criteria. Poor peak shape includes issues such as tailing, fronting, split peaks, and shouldered peaks.
The causes of poor peak shape are varied. They include everything from poorly developed methods, to contaminated columns or in other parts of the system, to extra system volume or poor system connections.
The following chart covers some of these common problems and possible remedies.
Problem
Indication
Cause
Solution
One or a few peaks tailing
Unexpected analyze interaction with solid support or stationary phase
All peaks tailing
Poor Connections in system
Void in Column
Use base deactivated, or hybrid silica materials. Old style silica have active sites on the surface that can have unintended interaction with sample analyze
Add a mobile phase modifier to the system such as TEA
Ensure that all tubing ends are cleanly and squarely cut.
Make sure that all nuts and ferrules are of the correct shape and tubing depth is correct for all fittings. Different manufactures have different tolerances, and they are not all interchangeable.
Use PEEK or universal spring loaded connecting tubing.
Replace Column
Initial Peaks fronting
Dirty guard Column
Replace Guard
Dirty column frit
Replace frit
All peaks fronting
Too strong a sample solvent
Samples should be dissolved in mobile phase or solvent weaker than mobile phase.
Column overload
Reduce injection volume
Reduce sample concentrations.
Split Peaks
Shoulder on 1 or 2 peaks
Shoulder on all peaks
Void or channel in column
Use a column with larger i.d.
Back flush or replace column
Dirty column frit
Replace frit
Dirty Column
Sample solvent not suitable for use with mobile phase
Back flush or replace column
Ensure that sample and mobile phase solvents are compatible make any necessary adjustments
Mobile phase is at 1 pH near the pKa of the analyze
Dirty Inlet frit
Dirty Guard Column
Dirty Column
Adjust pH of mobile phase to be at least 2 pH units away from the pKa of the sample analyzes
Adjust mobile phase to separate the 2 peaks
Replace frit
Replace Guard
Back flush or replace